GLD series features high perbance, low vibration and noise and several functions such as gas ballast valve, oil-back-flow prevention mechanism, and large sized oil level gauge. This series equips multi-voltage motor and correspondent to international standard.
Applications
•
Chemical, science experiment, Analyzer and Laser system.
Single phase, 100-120V/200-240V, 550W,
4 poles, Multiple-range motor, Capacitor start & run
Full load current
A
8.2 (100-120V)
4.1 (200-240V)
7.9 (100-120V)
3.9 (200-240V)
Oil capacity
mL
1,100
Recommended oil
SMR-100
Weight
kg
29
Inlet port diameter
mm
KF-25
Ambient temperature
°C
7 – 40
Overall dimensions
mm
170(W) x 515.5(L) x 249.5(H)
*Ultimate pressure is measured by Pirani gauge.
Caution
top of page
1.
Oil is deteriorated with operating, and becomes the factor of perbance shortage.
A decrease and the dirt of oil are confirmed, and please regularly add and replace it.
2.
A backflow preventer is incorporated into the pump to prevent the oil from flowing back while the pump is stopped.
Please make a suction side into atmospheric pressure when the pump is stopped for a long time.
3.
If operation is perbed at high pressure, oil mist is generated at the exhaust side.
4.
It is not suitable for the pump if operation is perbed continuously at a high evacuation pressure of 10kPa or more.
5.
There is a risk of explosion. Never block the outlet or operate the pump with equipment mounted at the outlet side that blocks the passage of gas.
Otherwise, the internal pump pressure increases causing the pump to explode, the oil level gauge to protrude or the motor to be overloaded.
6.
This motor houses an automatic returning thermal protector, which may be activated in case of trouble.
The motor will restart automatically when the temperature goes down.
If the thermal protector is activated, switch off the power immediaby for safety reason.
Outline Drawing
top of page
Pumping Speed Curves
top of page
Option
* You can see each option, if you click an option.