夜色网,色欲aⅴ精品一区二区三区四区,国产精品jizz视频国产y网,欧美精品综合视频一区二区

產(chǎn)品簡(jiǎn)介
TOTAL CARBON ANALYZER MODEL TCA08
TOTAL CARBON ANALYZER MODEL TCA08
產(chǎn)品價(jià)格:¥1
上架日期:2020-05-18 16:48:37
產(chǎn)地:美國(guó)
發(fā)貨地:美國(guó)
供應(yīng)數(shù)量:不限
最少起訂:1個(gè)
瀏覽量:193
資料下載:暫無(wú)資料下載
其他下載:暫無(wú)相關(guān)下載
詳細(xì)說(shuō)明

    供應(yīng)Magee黑碳儀

    深圳市現(xiàn)代豪方儀器儀表科技有限公司——來(lái)電請(qǐng)轉(zhuǎn) 余工

    TCA 08型總碳分析儀

    (TOTAL CARBON ANALYZER MODEL TCA08)

    PRODUCT INFO

    The analytical system of the Magee Scientific Total Carbon Analyzer is constructed entirely from stainless steel, making the instrument rugged, simple and reliable. The analysis uses ambient air as the carrier gas. This eliminates the need for specialty gas supplies, and greatly simplifies the installation, use, and maintenance of the instrument.

    In contrast to existing bs, the TCA offers greater reliability, greater flexibility, and substantial operational cost savings for aerosol analysis.

     

    The TCA08 can be used in online mode for continuous, real-time analysis: and offline mode to analyze previously-collected samples..

     

    ADVANTAGES AND BENEFITS

    CONTINUOUS OPERATION

    The timebase for sampling and analysis is adjustable from 20 minutes to 24 hours.

    NO GAS, NO GLASS

    Ambient air is used as the analytical carrier gas at a very low flow rate. This eliminates the need for specialized gas supplies.

    The analytical chambers are made of stainless steel: the instrument contains no fragile glass components.

    OC/EC DETERMINATION

    The Total Carbon Analyzer provides the result for TC.

    Since {TC = OC + EC}, it follows that {OC = TC EC}.

    The Total Carbon Analyzer may be connected to a Model AE33 Aethalometer, which provides real-time data for BC, closely related to EC. The Aethalometer data is automatically combined with the TCA data to provide a full characterization of the carbonaceous component of the aerosol in near-real-time: TC, OC, BC (or EC).

    CONTINUOUS OPERATION

    One channel collects the sample, while the second parallel channel is being analyzed. Ball valves automatically switch between the two, every timebase period, to provide continuous operation.

    QUICK INSTALLATION, AUTOMATIC OPERATION

    The instrument can be installed quickly with no specialized resources required. It operates automatically and recovers from power interruptions.

     

    OC/EC DETERMINATION

    Carbonaceous matter {TC = OC + EC} is often the largest contributor to PM2.5 mass. The Total Carbon Analyzer provides the result for TC.

    Since {TC = OC + EC}, it follows that {OC = TC EC}.

     

    The Magee Scientific Total Carbon Analyzer may be connected to a Model AE33 Aethalometer, which provides real-time data for BC, closely related to EC. The Aethalometer data is automatically combined with the TCA data to provide a full characterization of the carbonaceous component of the aerosol in near-real-time: TC, OC, BC (or EC).

    This is accomplished in a rugged, reliable instrument package that is suitable for laboratory, fieldwork, and routine Air Quality monitoring applications.

     

    RUGGED INSTRUMENT

    The Magee Scientific Total Carbon Analyzer Model TCA08 is constructed in a rugged instrument package suitable for laboratory, field-work and routine Air Quality monitoring applications.

     

    MEASUREMENT PRINCIPLE

    The Magee Scientific Total Carbon Analyzer collects a sample of atmospheric aerosols on a quartz fiber filter enclosed in a small stainless-steel chamber, at a controlled sampling flow rate of 16.7 LPM. The default sampling time is 60 minutes, but may be set from 20 minutes to 24 hours depending on the ambient aerosol concentrations.

    The instrument has two identical parallel channels, with the air flows being controlled by ball valves and solenoids. While one channel is collecting its sample, the other channel is analyzing the sample collected during the previous period. At the end of the period, the valves switch over to provide continuous operation and continuous data.

    After collection on the filter, two flash-heating elements combust the sample instantaneously in a small analyticflow of filtered ambient air. This converts all the carbonaceous compounds into CO2, and creates a short but large-amplitude pulse of CO2 in the analytic flow passed to the NDIR CO2 detector. The background level of CO2 in ambient air during the heating cycle is determined before and after the heating cycle, the provide the baselines against which the combustion pulse is measured. The CO2 concentration over the baseline is integrated to give the Total Carbon content of the sample.

    The chamber and combustion elements are cooled after analysis. At the end of the timebase period the flow system of ball valves and solenoids switches over. The first channel collects the next periods sample, while the second channel is analyzed.

     

     

    SPECIFICATIONS

    OPERATION

    Supply voltage: 100-240 V~, 50/60 Hz

    Max power consumption: 600 W (Heaters 400 W, Sample pump 100 W, Analytic flow pump 25 W)

    Airflow: Sample Airflow 16.7 LPM, Analytic Airflow 0.5 LPM

    Time resolution: 60 min (default); user-selectable 20 min to 24 hours

    ENVIRONMENTAL OPERATING CONDITIONS

    Indoor use, rack or benchtop (instrument)

    Altitude: up to 3000 m

    Temperature range: 10 40 degrees Celsius (instrument)

    Relative humidity range: non-condensing

    ANALYTICAL PERFORMANCE

    Sensitivity: 0.5 μg C/m3 at 60 min Sample time base and 16.7 LPM Sample Airflow

     

    MECHANICAL SPECIFICATION

    Chassis: standard 19’’/9U, rack mount, constructed of sheet bl

    Inlet plumbing requirement: the minimum overall height of the Total Carbon Analyzer (TCA) chassis and inlet tubing is 1.20 m. See Figure 1.

    Weight: 35 kg

     

    CONNECTORS

    Sampling air stream:

    l inlet: inner diameter 14 mm, outer diameter 18 mm, custom connectors. See Figure 2.

    outlet: .1/4NTPF

     

    Analytic air stream:

    l inlet: 1/8’’’ NTPF

    outlet: (internal, not brought to panel connector)

     

    Communications:

    l 6x COM, 4x USB 2.0, 1x USB 3.0, 3x power USB, Ethernet

     

    USER INTERFACE

    l 8.4SVGA display with LED backlight

    l Basic control: touch-screen

    l Optional control: standard PC keyboard and mouse

    l Red, Yellow, Green status LEDs

     

     

    深圳市現(xiàn)代豪方儀器儀表科技有限公司

    詢價(jià)請(qǐng)轉(zhuǎn)余工

在線詢盤(pán)/留言
  • 免責(zé)聲明:以上所展示的信息由企業(yè)自行提供,內(nèi)容的真實(shí)性、準(zhǔn)確性和合法性由發(fā)布企業(yè)負(fù)責(zé),本網(wǎng)對(duì)此不承擔(dān)任何保證責(zé)任。我們?cè)瓌t 上建議您選擇本網(wǎng)高級(jí)會(huì)員或VIP會(huì)員。
    企業(yè)信息
    深圳市現(xiàn)代豪方儀器儀表科技有限公司
    會(huì)員級(jí)別:試用會(huì)員
    ------------ 聯(lián)系方式 ------------
    聯(lián)系人:業(yè)務(wù)經(jīng)理(先生)
    聯(lián)系電話:0755-26998441
    聯(lián)系手機(jī):13246682860
    傳真號(hào)碼:0755-26738841
    企業(yè)郵箱:1540121144@qq.com
    網(wǎng)址:
    郵編:518057
    推薦供應(yīng)
    0571-87774297